It has been shown that, in addition to conventional contact electrode techniques, optical methods using fluorescent dyes can be successfully used for cardiac signal measurement. In this review, the physical and technical fundamentals of the method are described, as well as the properties of the most common systems for measuring action potentials and intracellular calcium concentration. Special attention is paid to summarizing limitations and trends in developing this method., M. Ronzhina, ... [et al.]., and Obsahuje seznam literatury
Motion artefact (MA) in voltage-sensitive fluorescent signals causes significant debasement of action potential. During ischemia and reperfusion in isolated rabbit heart, this artefact develops in a manner which may be described by the time of its onset, level, and shape. The MA during ischemia: (a) may become substantial with approximately two minutes delay after establishing global ischemia; (b) may be almost twice as high as the physiological action potential and decreases both with time and repetition of ischemia; (c) the MA shape is unpredictable and depends on individual rabbit., O. Janoušek, ... [et al.]., and Obsahuje seznam literatury